Ultraviolet Spectroscopy


| Home | Site Map | Scientific Staff | Related Links |


Agilent 8543 spectrophotometer


Procedure

    
The procedure for this section of the experiment is rather simple.  An Agilent 8543 spectrophotometer (pictured above) was used to measure the absorbance of our samples. Five-hundred
µL cuvette holds the sample being measured.  In order to blank the device, a cuvette is filled with 1X TE buffer.  The samples were loaded into the cuvette using a micro syringe.  The absorbance of each sample of green fluorescent protein was measured.  For sake of comparison, the absorbance of hemoglobin and lysozyme was taken.  The absorbance of each sample was then transferred onto a graph using computer analysis.  The final step is to determine the protein concentration from the absorbance, using Beer's Law.

Results

The protein concentration was determined using Beer's Law (A=abc), .where the extinction coefficient was 30,000 (M^-1)(cm^-1) and the path length = 1mm.   At 395nm, the green fluorescent protein  had an absorbance of 2.8028x10^-3, which gives it an concentration (c) of 9.34x10^-7.  A second reading was taken to insure accuracy in our measurement.  Also, at 395nm, the absorbance was 2.8596x10^-3, which gives concentration (c) of  9.532x10^-7.